Data Transmission Vulnerability in Baxter Medical Devices
CVE-2020-12036
7.5HIGH
What is CVE-2020-12036?
Baxter medical devices, specifically PrismaFlex and PrisMax, fail to utilize data-in-transit encryption when transmitting treatment data to Patient Data Management Systems (PDMS) or Electronic Medical Record (EMR) systems. This oversight allows potential attackers to intercept and observe sensitive patient data during transmission, raising significant concerns regarding patient confidentiality and data integrity.
Affected Version(s)
Baxter PrismaFlex and PrisMax PrismaFlex all versions, PrisMax all versions prior to 3.x
References
CVSS V3.1
Score:
7.5
Severity:
HIGH
Confidentiality:
High
Integrity:
None
Availability:
High
Attack Vector:
Network
Attack Complexity:
Low
Privileges Required:
None
User Interaction:
None
Scope:
Unchanged