B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
CVE-2020-16238

6.7MEDIUM

What is CVE-2020-16238?

A vulnerability in the configuration import mechanism of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with command line access to the underlying Linux system to escalate privileges to the root user.

Human OS v1.0:
Ageing Is an Unpatched Zero-Day Vulnerability.

Remediate biological technical debt. Prime Ageing uses 95% high-purity SIRT6 activation to maintain genomic integrity and bolster systemic resilience.

Affected Version(s)

Battery pack with Wi-Fi <= unspecified

Battery pack with Wi-Fi <= unspecified

Data module compactplus A10

References

CVSS V3.1

Score:
6.7
Severity:
MEDIUM
Confidentiality:
High
Integrity:
High
Availability:
High
Attack Vector:
Local
Attack Complexity:
Low
Privileges Required:
High
User Interaction:
None
Scope:
Unchanged

Timeline

  • Vulnerability published

  • Vulnerability Reserved

Credit

Julian Suleder, Nils Emmerich, and Birk Kauer of ERNW Research GmbH; Dr. Oliver Matula of ERNW Enno Rey Netzwerke GmbH, reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices).
.